5400 Series Semi-Automatic Manifold Systems
5400 Series Semi-Automatic Manifold Systems
5400 series semi-automatic manifold system is designed to provide an uninterrupted gas supply. It consists of a primary bank and a reserve bank of cylinders. When the pressure in the primary cylinder bank reduces to the pre-set value, the changeover takes place automatically to provide continuous supply of gas from the reserve bank. Upon changing the cylinders, the regulators on both banks need to be re-adjusted in order for the changeover to occur automatically next time. The secondary regulator in the main pipeline stabilizes the outlet gas flow.
Features
Open-style manifold
Secondary regulator for consistant high flow delivery pressure to the pipeline
Silver brazing on piping joints for maximum leak prevention
System is designed to accommodate future expanison needs
System is mounted with gas filters
Unique changeover valve provides uninterrupted supply of gas from primary and reserve banks
Pressure switch port is available
Headers have been tested to withstand high cylinder pressure
Wall or floor mount available
Installation Dimensions
Installation Dimensions
Manifold System Layouts
Manifold System Layouts
Ordering Information
Gas Service Series Max. Inlet Pressure Delivery Pressure Max. Delivery Flow Outlet Connection Pigtail Specifications
Oxygen
5300X
5300XH
2175 psi (150 bar)
3000 psi (207 bar)
10~200 psi (0.7~14 bar)
10~200 psi (0.7~14 bar)
3500 SCFH (100 m3/h)
3500 SCFH (100 m3/h)
3/4" NPT (M)
3/4" NPT (M)
Pigtail, CGA540
Pigtail, CGA540
Carbon Dioxide
5300C
3000 psi (207 bar)
5~125 psi (0.35~8.6 bar)
2100 SCFH (60 m3/h)
3/4" NPT (M)
Pigtail, CGA320
Argon
5300IN
5300INH
2175 psi (150 bar)
3000 psi (207 bar)
10~200 psi (0.7~14 bar)
10~200 psi (0.7~14 bar)
1750 SCFH (50 m3/h)
1750 SCFH (50 m3/h)
3/4" NPT (M)
3/4" NPT (M)
Pigtail, CGA580
Pigtail, CGA580
Nitrogen
5300IN
5300INH
2175 psi (150 bar)
3000 psi (207 bar)
10~200 psi (0.7~14 bar)
10~200 psi (0.7~14 bar)
7000 SCFH (200 m3/h)
7000 SCFH (200 m3/h)
3/4" NPT (M)
3/4" NPT (M)
Pigtail, CGA580
Pigtail, CGA580
Helium
5300IN
5300INH
2175 psi (150 bar)
3000 psi (207 bar)
10~200 psi (0.7~14 bar)
10~200 psi (0.7~14 bar)
3500 SCFH (100 m3/h)
3500 SCFH (100 m3/h)
3/4" NPT (M)
3/4" NPT (M)
Pigtail, CGA580
Pigtail, CGA580
Air
5300Q
5300QH
2175 psi (150 bar)
3000 psi (207 bar)
10~200 psi (0.7~14 bar)
10~200 psi (0.7~14 bar)
3500 SCFH (100 m3/h)
3500 SCFH (100 m3/h)
3/4" NPT (M)
3/4" NPT (M)
Pigtail, CGA346
Pigtail, CGA346
54
Series
1
Manifold System Layout
2
Cylinder Valve Spacing
X
Gas Service
H
Max. Inlet Pressure
- 5 x 5
Number of Cylinders (left-hand / right-hand)
-1
Type of Mounting
54 1: Standard layout
2: “L"" Shape layout
3: “U"" shape layout
4: Crossover layout
5: Staggered layout
1: 5"" (127 mm)
2: 10"" (254 mm)
3: 13"" (330 mm)
4: 18"" (457 mm)
X: Oxygen
C: Carbon Dioxide
IN: Ar, He, N2
Q: Air
None: 2175 psi
H: 3000 psi
1 x 2: One cylinder on the left,
Two cylinders on the Right
5 x 5: Five cylinders on the left,
Five cylinders on the Right
...
Note: Direction of piping (Right or Left) is
indicated by facing the manifold.
1: Wall mount
2: Floor mount
Download